      Figure 1.11b represents a setup of PL excitation and emission spectra by using the integration sphere. Generally, PL can be observed upon UV–VIS excitation from Xe‐lamp, and we can select which luminescence center (or electron transition) can be induced by selecting the excitation wavelength using a grating. If we use the integration sphere, we can collect all the PL photons. If we define A and I as absorption, which is evaluated by the intensity of excitation light with and without the sample and PL emission intensity, respectively, we can deduce Q = I/A experimentally. The detectors for PL photons are generally grating and photodetector. PL quantum yield is a quantitative value, but generally PL is a qualitative evaluation with an arbitrary unit, since PL intensity depends on the geometry of sample setting. In common PL measurement, we do not use the integration sphere.

      (1.71)equation

      (1.72)equation

      (1.73)equation

      (1.74)equation

      where kf and knr denote rate constants of radiative and non‐radiative transitions, respectively. In this way, we can evaluate radiative and non‐radiative rate constants quantitatively.

Schematic illustration of PL decay setup. Schematic illustration of the common setup of X-ray induced radio luminescence spectrum measurement. Schematic illustration of the common setup of gamma-ray induced scintillation decay curve measurement.

      Although DCM with a 511 keV γ‐ray source is the most common way to evaluate scintillation decay time, it contains several technical disadvantages. Because the energy of γ‐ray (511 keV) is high and has a high penetrative power of materials, detection efficiency is not high and requires a long time for the measurement to be made. In addition, low detection efficiency makes it difficult to measure a slow component, and generally, a component slower than several μs is difficult to measure. For example, measurements of emissions from Eu3+, Tb3+ and most transition metal ions, are almost impossible, although they are used for integration‐type detectors. The lack of the wavelength resolution is also a problem because identification of the scintillation emission origin is sometimes difficult. In most cases, we can guess the emission origin by PL decay, but decays of

Скачать книгу