ТОП просматриваемых книг сайта:
Body Sensor Networking, Design and Algorithms. Saeid Sanei
Читать онлайн.Название Body Sensor Networking, Design and Algorithms
Год выпуска 0
isbn 9781119390015
Автор произведения Saeid Sanei
Жанр Программы
Издательство John Wiley & Sons Limited
Figure 3.5 (a) ECoG and (b) foramen ovale electrodes denoted by pointers. In this setup, a scalp EEG has also been used. (See color plate section for color representation of this figure)
One of the challenging problems in assessing the recorded stethoscope data is separation of heart and lung sounds for automatic diagnosis. The difficulty stems from the fact that sound signals are subject to delay and for their separation a robust convolutive source separation technique is required. Although many solutions have been proposed by signal processing researchers [37–40], design of a clinically usable system which can effectively separate these two sound signals is still being researched.
Figure 3.6 New patch type ECG systems including electrodes and wireless connection.
3.3.2 Other Sensors
Oximeters are used to measure the blood oxygen saturation level (SpO2). The traditional measure of SpO2, called photoplethysmography, involves using two light beams in red and infrared wavelengths each sensitive to high and low blood oxygen levels respectively.
In principle, oximeters can operate in either transmissive or reflective mode. Those using an earlobe or finger measure the absorbed transmitted light through the tissue and others such as those used on the wrist work based on the reflective light by the tissue layers including the blood.
Figure 3.7 Electronic stethoscope; Littmann 3200 model.
In general, the SpO2 value is linearly dependent on the ratio between the transmitted infrared to red lights and decreases with this ratio. Figure 3.8 shows oximeters used in different modes of oximetry.
It is logical to accept that the oximeter reading is also an indicator of heart rate as the blood oxygen changes following the heart pumping cycle. Figure 3.9 shows two different oximeters with different recording modalities. The most important application of oximeters is in hospital intensive care units (ICUs) to ensure that the patient has normal breathing and heartbeat.
Amongst the other physiological measures, respiratory rate (RR) is important for the diagnosis of lung diseases and many other internal abnormalities. Therefore, it has an important role in healthcare and sport. In hospitals it is measured simply by the manual counting of breaths, conventional impedance pneumography, or an optical technique called end-tidal carbon dioxide (et-CO2).
Figure 3.8 Pictorial illustration of the concepts of transmissive and reflective oximetry.
Figure 3.9 Two different models of oximeters: (a) digital pulse oximeter used on the finger and (b) wireless wrist pulse oximeter, both from Viatom.
The impedance pneumograph is a bioimpedance recorder for indirect measurement of respiration. Using superficial thoracic electrodes, the system measures respiratory volume and rate through the relationship between respiratory depth and thoracic impedance change [41].
In the et-CO2 system, a device called a capnometer uses a small plastic tube inserted in the patient's mouth. The capnometer tube accumulates the expelled CO2 in each breath. An infrared light is then emitted to the CO2 from a moving source. Following absorption of infrared radiation overtime, a variation with breathing rate will be induced. The et-CO2-based systems are used in hospitals mainly for patients in critical conditions often within ICUs as well as ambulances for emergency checking of patient states. The capnographs, which look approximately like square waveform in normal conditions, significantly change during intubation, after cardiac arrest, ventilation problem, shock, emphysema, or leaking alveoli in pneumothorax, hypoxia due to asthma or mechanical obstruction, poor lung compliance, obese, and pregnancy [42].
Spirometry used for respiration monitoring is also one of the most common lung function tests. It measures the amount of air one can inhale and exhale. It also measures how fast the air can be emptied out of lungs. It is used to help diagnose breathing problems such as asthma and chronic obstructive pulmonary disease (COPD). During the test, the subject breathes in as much air as they can and then quickly blows as much air out as possible through a tube connected to a machine called a spirometer.
Body plethysmography is another common lung function test. It measures how much air is actually in the lungs when one inhales deeply. It also checks how much air remains in the lungs after they breathe out as much as they can. A plethysmograph has the same applications as a spirometer.
3.4 Biological Sensors
The need for biological sensors was perhaps initially raised by the Department of Defense in the United States. In 2004 they announced an urgent need for the development of sensors for early-warning systems and the protection of military forces against potential chemical and biological attacks [43]. Traditionally, these sensors are particular types of chemical sensors and benefit from the high selectivity and sensitivity of biologically active materials.
Such sensors produce an electrical signal proportional to the concentration of a specific chemical or set of chemicals in the human body. Therefore, they should be biocompatible with and have no toxic effect on humans. However, current biosensors are more than a chemical sensor and are often used in biomedical applications.
The biosensors generally consist of a transducer, biological-recognition membrane in intimate contact with the transducer, and biologically active material, which is sensitive to the analyte molecule through a shape-specific recognition. Biosensors are generally divided into two categories: affinity-based and metabolic.
3.4.1 The Structures of Biological Sensors – The Principles
A biosensor consists of a biological sensing element and a transducer. The sensing elements can be organisms, tissues, cells, organelles, membranes, enzymes, receptors, antibodies, or nucleic acids. On the other hand, transducer types can be electrochemical, optical, calorimetric (thermal), or acoustic.
Bio-affinity transducers rely on very strong binding through which the transducer detects the bound receptor-analyte pair. The most common bio-affinity recognition processes are receptor-ligand binding and antibody-antigen binding.
Metabolic biosensors, on the other hand, sense the instances where the analytes are bound or the co-reactants are chemically altered, which in turn product molecules are formed, and the transducer detects the changes in concentration of the product molecules or co-reactants and consequently heat is released by the reaction. The common biometabolic processes are enzyme-substrate reactions and metabolism of specific molecules by