ТОП просматриваемых книг сайта:
Process Gas Chromatographs. Tony Waters
Читать онлайн.Название Process Gas Chromatographs
Год выпуска 0
isbn 9781119633013
Автор произведения Tony Waters
Жанр Отраслевые издания
Издательство John Wiley & Sons Limited
SCI-FILE: On Column Types
Introduction to SCI‐FILEs
The more theoretical and mathematical content of the book resides in separate segments called SCI‐FILEs. These contain optional reading that may or may not be part of a course of study.
Each SCI‐FILE is a supplement to the main text that you can safely omit if not of immediate interest. Treat them as reference sources to consult when needed.
Two kinds of column
The stationary phase must be secure inside the column so it doesn't move. The packed column and the open‐tubular column differ per the method they use to anchor the stationary phase in place.
Packed columns
A packed column most often uses several meters of ⅛‐inch o.d. stainless steel tubing, although early PGCs used larger diameters, and some PGCs now employ
In the traditional packed column, the packing is a granular porous solid with particles about the same size as granulated sugar. These particles pack tightly together inside the tube so that any sample molecules moving with the carrier gas are in intimate contact with them.
The type of column so produced depends on the role of the solid particles:
An active‐solid column contains solid particles having a large activated surface area to selectively adsorb certain molecules from the sample gas.Since the stationary phase is solid, this technique is gas‐solid chromatography (GSC).
A liquid‐phase column contains solid particles having a coating of non‐volatile liquid to selectively dissolve certain molecules from the sample gas.Since the stationary phase is liquid, this technique is gas‐liquid chromatography (GLC).
Many columns now use proprietary stationary phases, often made from specialized polymer material. These columns don't easily fit into the old classifications of GSC or GLC, so the terminology is becoming passé.
In a liquid‐phase column, we call the granular solid an inert support. In real life, an inert support might not be completely inert; it sometimes affects the performance of a column.
The thickness of the liquid film coated on the support is an important variable. The liquid loading gives the percentage by weight of liquid on support.
The first gas chromatographs used packed columns, and they are commonly found in PGCs today.
Open‐tubular columns
An open‐tubular or capillary column uses several tens of meters of capillary tubing having an internal diameter ranging from about 100 to 530 μm. The mode of operation differs: the stationary phase adheres to the inner wall of the tube, and the carrier gas flows down the middle. Figure 1.6 illustrates three versions (Harvey 2017):
A “wall‐coated open‐tubular” or WCOT column uses tubing made of fused silica. The stationary phase is a very thin layer of a non‐volatile liquid coated on the inside wall of the tube to selectively dissolve sample molecules from the sample gas. PGCs rarely use these columns as they are fragile and tend to be unstable in use.The WCOT columns typically use fused‐silica tubing and tend to be too fragile for process use. The PLOT and SCOT columns mostly use steel capillary tubing. Figure 1.6 Three Kinds of Capillary Column.
A “porous‐layer open‐tubular” or PLOT column uses stainless steel capillary tubing. The stationary phase is a very thin layer of solid material coated on the inside wall of the tube to selectively adsorb sample molecules from the sample gas.
A “support‐coated open‐tubular” or SCOT column typically uses stainless steel capillarytubing. The stationary phase is a coating on very fine support particles in a uniform layer on the inner wall of the tube. These rugged columns have become quite popular in process gas chromatographs.
Open‐tubular columns have smaller diameters than packed columns and require special operating techniques. As in packed columns, the film thickness is an important variable, but we'll defer discussion on that. While they achieve better separations, the operating conditions of open‐tubular columns can be difficult to sustain in the process environment.
For more information about column types and column liquid phases, refer to the excellent detailed review by Rahman et al. (2015).
The detector
Making the measurements
A chromatographic separation cannot produce a measurement. Chromatography is merely a separating technique; it doesn't measure anything. To measure the concentration of the analytes, the analytical instrument must estimate the quantity of selected molecules as they elute from the column. It follows that every gas chromatograph needs a device to generate a signal proportional to the number of sample molecules exiting the column. This is what a detector does.
In any gas chromatograph, two things are happening in series. First the column separates the analytes, and then the detector measures them. To improve your troubleshooting ability, keep that distinction in mind.
Many detectors are available for gas chromatography, most developed for applications that require selective measurement or enhanced sensitivity.
The thermal conductivity detector (TCD) was the first gas chromatograph detector, and after much improvement it is still popular today. The TCD responds to the difference in thermal conductivity between pure carrier gas and carrier gas that contains sample molecules. So, when a TCD is used, the carrier gas is chosen to maximize the difference in thermal conductivity between the carrier gas and the analytes. The TCD is a general‐purpose detector that will respond to any analyte.
Most other detectors are selective; they respond only to certain kinds of molecules and often do so with very high sensitivity. For instance, the flame ionization detector (FID) responds only to compounds containing both carbon and hydrogen, so it's very useful in the analysis of hydrocarbons. The flame photometric detector (FPD) is also very sensitive, but only to sulfur or phosphorus compounds. It's most used to measure sulfur compounds in fuels and stack emissions to ensure compliance with environmental regulations.
Generally, detectors operate in the differential mode. When pure carrier gas is passing through a detector, its output signal should be constant. The analytical instrument reads that signal and offsets it to a value close to zero. We call that the baseline. Then, when the detector responds to the presence of analyte molecules, the instrument outputs a change in signal level proportional to the concentration of that component.
Chapter 10 provides a detailed review of the three detectors most used in process gas chromatographs (TCD, FID, and FPD) and briefly mentions some other detectors that are common in laboratory instruments but only occasionally deployed for online process applications.
The chromatogram
The chromatogram is a graphical display of the detector signal plotted against elapsed time. The PGC may print the chromatogram on a chart or display it on a computer screen. Note that it's also common to refer to the raw signal from a detector as the chromatogram signal.