Скачать книгу

many situations in which these differences are small. In this case, a reduced contrast NMR image will be recorded. Usually, even for the same tissue, there can be differences in the values of T1 and T2 relaxation times; T1 being usually longer than T2. As such, it becomes possible to record either T1 or T2 weighted MRI images. The T1‐weighted NMR images are obtained by reducing the repetition time (i.e. the time between two successive pulses), while the T2‐weighted MRI images are recorded by reducing the echo time (the time after which the magnetic resonance signal is recorded).

      Over time, it was demonstrated that the contrast of MRI images can be increased by using different contrast agents (CAs). The first CAs that have been employed are the paramagnetic ions possessing large magnetic momentum (Mn2+, Gd3+). Their effect on the effective relaxation time is measured in terms of relaxivity, according to the following equation:

equation

      where i is either 1 or 2, and Ri,obs is the observed relaxation rate which is defined as the reverse of the effective relaxation time Ti,eff. Ti is the relaxation time in the absence of any CA, c is the CA concentration, and ri is the relaxivity, which is characteristic of CA. Higher the ri values are, higher the effectiveness of the CA, meaning that a smaller concentration of CA leads to measurable effects. The effects of the CA on the two relaxivities are very often completely different. As such, CAs that reduce the T1 values (high r1) and increase the signal in T1‐weighted images are called positive CAs, whereas CAs that reduce the T2 values and subsequently reduce the intensity of the signal in T2‐weighted images are called negative CAs.

      The mechanism by which CAs influence the relaxation times is related to the value of their magnetic moment and the time spent by the protons (mostly belonging to water molecules) in their proximity (retention time). Thus, paramagnetic ions with large magnetic moments (Mn2+, Gd3+, Dy3+) allow a close interaction of the ions with water molecules present in the first hydration sphere, leading to an increase of r1 relaxivity values. These paramagnetic ions are usually administrated in the form of ion complexes that have the role to decrease their mobility and to increase their relaxivities. The phenomenon is known as Proton Relaxation Enhancement (PRE). These ions are known as T1‐positive CAs.

      On the other hand, the advancement in the synthesis of different classes of MNPs with improved magnetic properties has represented a major discovery in the field of CAs. MNPs have huge values of their magnetic moments, with respect to paramagnetic ions, influencing the relaxation behavior of water molecules beyond the first hydration layer. Nevertheless, their coatings, used for improving their biocompatibility, impede a direct contact with water molecules. For these reasons, the MNPs are mostly used as negative T2‐CAs. The first commercially available MNPs approved for clinical applications as CAs in MRI are Resovist and Feridex (Table 2.1). They consist of SPIONs of small magnetic cores (approximately ~10 nm) obtained by the coprecipitation method. Their r2 values are lower than 100 s−1 mM−1. The relaxivity values are directly proportional to the square of MNPs magnetic moment, which in turn is proportional to their magnetization saturation (Ms) and volume. Therefore, MNPs with larger volumes and Ms values present several fold higher relaxivities. The use of thermal decomposition methods led to the synthesis of MNPs with improved crystallinity. This increases their magnetic moments as well as their relaxivities. However, monocrystalline MNPs with sizes higher than 20 nm are difficult to obtain because they involve laborious seed‐mediated growth synthesis methods. Moreover, the increase of MNP size can induce their transition into the so‐called “blocked state” in which they will have a nonzero remanent magnetization (Mr – induced magnetization remaining at zero magnetic field). A direct consequence of this nonzero Mr will be the occurrence of dipolar interaction and MNPs' self‐aggregation that make them less suitable for biomedical applications. For example, a 20 nm diameter appears to be an upper limit in the case of MNPs applications such as MRI CAs.

      Interestingly, in the case of longitudinal relaxivity, it has been observed that its value tends to decrease with increasing the size of the cluster. This effect is explained by the reduction of the surface accessible for water molecules as the size of the cluster increases.

Schematic illustration of example of a NMRD profile for a colloidal suspension of SPIONs showing the evolution of the r1 proton relaxivity with the applied external magnetic field/frequency.

      Source: Reprinted with permission from Laurent et al. (2008). Copyright 2019 American Chemical Society.

      By fitting the experimental data with adequate theories, various parameters can be extracted, such as Rmax, the maximal relaxivity, Ms of the MNPs, tD, water diffusion correlation time, and MNPs' anisotropy energy (Muller et al. 2005).

      But the most common application of MNPs, like MRI CAs, is the early detection and diagnosis of cancer. The MNPs' effectiveness in cancer–tumor detection is mainly dictated by their fate after they are administrated in the systemic circulation.

Скачать книгу